Small Thin Film Analyzer
Specifications
Sample Sizes
| Standard
load based test |
3.08cm
x 15.3cm (1.21" x 6.02") By special order (3.08cm X 17.78cm) |
| High
load test |
May
require a necked (dog bone) sample, depending on expected load: necked
down to 1.50 cm (0.59") |
| Minimum
thickness |
The
actual material properties will determine the minimum thickness. |
| Maximum
thickness |
3mm
(0.12"), depending on clamp foot design |
| Short
column test (Option) |
5.1cm
X 3.2cm (2" x 1.25".) This sample size may vary depending on
ultimate load potential, flute type etc. |
| Option
1 |
Pass
through sample preparation punches |
Load Range
| Internal
load frame capacity |
±1.3344kN
(300 lbf) |
| A/D
conversion |
16
bits resolution |
| Load
control |
16
bits resolution. NOTE: this device's cross head motion uses load control. |
| Load
ramp rates |
User
selected, computer generated. |
| Calibration |
444.8
N (100 lbf) load ring standard |
| Option
1 |
Short
column test fixtures. |
Extensometer
| A
precision extensometer is used for motion sensing. Typical setups use
25.4mm (1") gage length 10% extensometer. A/D conversion is at 16 bit
resolution. |
| Option
1 |
25.4mm
(1") gage extender for 5% operations |
| Option
2 |
1 -
2mm (micron level) measurement |
| Option
3 |
Optical
measurements |
Software
| Option
1 |
Microsoft®
SQL Server® data storage |
Relative
Humidity
| The
standard setup is in a 22.8°C (73°F) 50% room. The system can be placed
in other environmental conditions, such as a 20% room or a 90% room.
Please contact us for details and suggestions. |
| Option
1 |
Below
10% operations |
| Option
2 |
Above
95% operations - Condensing water may result in erratic results. |
Temperature
Control
| The
standard setup exchanges heat to ambient room conditions. |
| Standard |
22.8°C
(73°F); 50% RH. |
Connections
| Power |
Typically
120VAC 1 phase 10A 60Hz |
| Auxiliary
inputs |
Room
condition recording. |
| Vacuum
pump exhaust vent |
Vented
outside preferable. |
| Option
1 |
House
vacuum if available, may be used. |
| Option
2 |
Analog
phone line |
| Option
3 |
Network
connection |
| Option
4 |
Printer |
| Option
5 |
Alarms |
Dimensions
| Actual
placement requires extra room around the units |
| Test
unit |
45.72cm
L x 68.58cm W x 22.86cm H (18"L x 27"W x 9"H) |
| Vacuum cart |
50.8cm
L x 101.6cm W x 50.8cm H (20"L x 40"W x 20"H) |
System configuration and specifications may
change without notice.
Delphinus reserves the right to make, from time to time, such departures from
the detail specifications as may be required to permit improvements in design of
its products. Components made under military approvals will be in accordance
with the approval requirements. The information included herein is believed to
be accurate and reliable. However, Delphinus assumes no responsibility for its
use; nor for any infringements of patents or other rights of third parties that
may result from its use.
|
|
|
|
|
|
|
|