Altitude Control Chamber Specifications

Chamber Sizes

Standard Chamber By special order, such as 60.96cm x 60.96cm x 60.96cm (24" x 24" x 24") 
Chamber size ranges 1049cu cm to 1812278.18cu cm (64cu " to 110592 cu ")
Door Sizes  Application will determine
View ports Application will determine

Environmental Range (an example)

Temperature Maximum 93C (200F) (Typical)
Temperature Minimum -56C (-70F) (Typical)
Temperature Change Rate 30C / minute (both directions) (Typical)
RH Minimum Near 0%RH dependant on temperature and pressure
RH Maximum Near 100% dependant on temperature and pressure
RH Change Rate 20% / minute (both directions) (Typical)
Pressure Maximum 1.15 Bars (1.136atm) (862.5 torr) (33.96"Hg)
Pressure Minimum 50 mBars (0.0194 atm) (37.5 torr) (1.476"Hg)
Pressure Change Rate Pressure 800 torr/ min. Vacuum 760 torr to 100 torr 1 minute. 100 to 37.5 in 5 minutes

Chamber Construction

Test chamber constructed out of stainless steel.  Various finishes are available. Chamber heat exchange can be provided by outer wall (external) or a heat plate (internal).
Option 1 Air Circulation fans.
Option 2 Electronics feed-throughs
Option 3 ESD protection

Software

Option 1 Microsoft® SQL Server® data storage

Relative Humidity

The standard setup is in a 22.8°C (73°F) room. Please contact us for details and suggestions.
Option 1 Below 10% operations
Option 2 Above 95% operations - Condensing water may result in erratic results.

Temperature Control

The standard setup exchanges heat to ambient room conditions using two stage chillers and electric heaters.
Working Fluid Dependant on conditions and chamber design

Connections

Power Typically 240VAC 3 phase 50A 60Hz
Auxiliary inputs Room condition recording.
Vacuum pumps exhaust vent Vented outside required. 
Option 1 House vacuum if available, may be used.
Nitrogen Supply Nitrogen Purge
Air pressure House Air Pressure may be used.
Distilled water Distilled water systems
Water drain Water drain systems
Option 2 Analog phone line
Option 3 Network connection
Option 4 Printer
Option 5 Alarms

Dimensions

Actual placement requires extra room around the units
Test unit Dependant on requirements
Utility cart Dependant on requirements

System configuration and specifications may change without notice.

Delphinus reserves the right to make, from time to time, such departures from the detail specifications as may be required to permit improvements in design of its products.  Components made under military approvals will be in accordance with the approval requirements.  The information included herein is believed to be accurate and reliable.  However, Delphinus assumes no responsibility for its use; nor for any infringements of patents or other rights of third parties that may result from its use.

Free Tips:

Microsoft Windows 7 migrations can be implemented quicker by using the old system computers with a Remote Desktop Connection. (01/19/11)

(More tips)

 

News:

(News)

 

Send us a question

Submit technical question.
Top

Send questions and comments about this website to: contact@delphinus-llc.com
© 2011 Delphinus, LLC.  All rights reserved.
 
           
Home Computers Software development Test equipment Contact us About us